FRONT PANEL OPERATION
Continuous Transient Operation
In this mode, electronic load will generates a repetitive pulse stream that toggles between two
load levels. ; Load could switch the state between two setting value (value A and value B).
In this following example, assume that the CC mode is active; the applicable transient operation
parameters have been set as follows.
For example:
Continuous mode, current level A =5A, width = 2mS. Current level B =10A, width = 3ms. Testing
machine input voltage is 12V.
10A
5A
2.0ms 3.0ms
Continuous Transient Operation
Action
1. Select the operation mode (CC,CV,CR or CP), and set a proper value, press On/Off to
turn off the load input.
2. Press Shift + S-Tran , set LEVEL A=5A, press Enter , set WIDTH A=3mS, press
Enter set LEVER B=10A, press Enter , set WIDTH B=2mS, press Enter to confirm.
3. Transition mode now is CONTINOUS, press Enter to confirm.
4. Press Shift + Tran to activate the transient mode.
5. Press Shift + Tran again to stop the transient operation.
6. Press On/Off , the input of load is powered on.
Pulse Transient Operation
In this mode, generates a transient pulse of programmable width when pulsed transient
operation is in effect.
For example: When load receiving one trigger signal, it will switch to 10A current value, and
taking 10mS to return the current value of 5A.