FRONT PANEL OPERATION
Action
1. Press Shift + S-Tran , Setting LEVER A=5A, LEVER B=10A, transition mode is TOGGLED.
2. Press Shift + Tran to activate the transient mode.
3. Press Shift + Trigger switch to the current value of 10A.
4. Press Shift + Trigger switch between 5A and 10A.
5. Press Shift + Tran again to stop the transient operation.
List Operation
Users can use the front panel keypad or Power View 8500(PV8500) software to
programming the list sequence. Please refer to the software user’s guide.
The following example will help you how to do the list operation in front panel.
Trigger
0
1 2 34 5
List count=1
List count=2
List sequence
Action
1) Press the button of On/Off , execute the input of Load I in OFF state.
2) Press Shift + Menu , move cursor to the option of menu of CONFIG, Press Enter
into the next step menu, move cursor to TRIGGER SOURCE. PresEsnter
and move
cursor to IMMEDIATE <DEF>, setting trigger source mode is panel IMMEDIATE mode.
3) Press Enter to confirm.
4) Press Esc to the previous menu, move cursor to LIST SET.
5) Press Enter into the next step menu .move cursor to EDIT LIST FILE.
6) Press Enter into the next step menu, move cursor to CURRENT LIST, select
CURRENT MODE.
7) Press Enter , move cursor to REPEAT. Setup LIST is in cycle mode.
8) Press Enter to confirm, setup the list steps = 5.
9) Press Enter to confirm, setup step 1 current =3A.
USER’S GUIDE 35