FRONT PANEL OPERATION
10A
5A
TWD
10ms
TWD
10ms
TRIG
TRIG
Pulsed Transient Operation
Action
1. Select the operation mode (CC,CV,CR or CP), and set a proper value, press On/Off to
turn off the load input.
2. Press Shift + S-Tran , set LEVER A=5A, LEVER B=10A, WIDTH B=10mS. Transition
mode now is CONTINOUS.
3. Press ▼ till >PULSE.
4. Press Enter , set transient mode is PULSE.
5. Press Shift + Tran to activate the transient mode.
6. Press Shift + Trigger to start another pulse. Press Shift + Trigger , get more pulse.
7. Press Shift + Tran again to stop the transient operation.
8. Press On/Off , the input of load is powered on.
Toggled Transient Operation
In this mode, after transition operation start, Load could change the input between the main
level and the transient level when toggled transient operation is in effect.
For example:
When Load receives one trigger signal, Load current will switch between 5A and 10A.
10A
5A
TRG
TRG
Toggled Transient Operation
34 USER’S GUIDE