FRONT PANEL OPERATION
Action
10) Press Enter to confirm, setup step 1 width=6mS.
11) Repeat 7) and 8) operation , set current and width of one step 0A, 5mS;2A,
4mS;6A,2mS;0A,5mS.
12) Press Enter to confirm, Menu STORE LIST FILE 1, save file in group1.
13) Move cursor to Mode Set , press Enter to enter into the next step menu to set mode
is <LIST MODE >
14) Press Enter to confirm
15) Press Esc , Press On/Off , make Load in ON state. Press Shift + Trigger , make list
operation run or stop.
16) Stop the list operation mode. Move cursor to LIST SET, press Enter , move cursor to
Mode Set in option menu, press Enter , enter into next step menu. Select mode is <FIXED
MODE>.
17) Press Enter to confirm.
Fast Recall Function
You recall 10 groups numbers very conveniently which were stored before.
Action
1) Press Shift + Menu
2) Press▼to CONFIG, press Enter into submenu, press▼ to SHORTCURT RECALL
3) Press Enter , select >ON and press Enter to confirm.
4) Press Esc two times and quit the menu.
5) Press 1 and you can get the first group number. Press 2 , you can get the second
group number. Press 0 , you can get the tenth group number. And if there is not stored
number, VFD will show NO EEPROM DATA.
Automatic Testing Function of Electronic Load
You test various parameters of power supplies at different loads. Now we take a kind of
charger as an example to explain the usage of the automatic testing of the electronic load.
36 USER’S GUIDE