Qdatasheet_Logo
Integrated circuits, Transistor, Semiconductors Search and Datasheet PDF Download Site

IT8518C View Datasheet(PDF) - ITE Tech. INC.

Part Name
Description
MFG CO.
'IT8518C' PDF : 44 Pages View PDF
FRONT PANEL OPERATION
process
1st step
2nd step
3rd step
4th step
5th step
6th step
The test proceeding of R35 charger
Method
Mode
Voltage(V) Current(mA)
Power(W)
CC
5.86.15
210
≤4(input175VAC)
None
5.96.4
0
≤1.2(input175VAC)
CC
5
205245
CV
3
205245
CV
2
205245
Short circuit
0
≤245
Ripple counter
≤ 50mVpp
Edit the Testing File
1) Press Shift + Menu into menu, VFD displays CONFIG
2) Press▼ to LIST SET.
3) Press Enter into submenu, press▼ toEDIT LIST FILE.
4) Press Enter and begin to edit the automatic testing file, VFD shows MAX CURR=
20.000A, which means setting the maximum of voltage. The maximum is bigger than 3A,
which means CC mode is in high range, here, the maximum is 3A. Press 3 and press Enter
to confirm.
5) VFD displays MAX VOLT=120.00V, which means setting the maximum of voltage, the
maximum is bigger than 18V, which means CV is in high range, here ,the maximum is 18V.
Press 1 + 8 to set MAX VOLT=18V, press Enter to confirm.
6) VFD displays MAX POWER=200.00W, which means setting the maximum of power,
here, it is150W, Press 1 + 5 + 0 to set MAX POWER=150W, press Enter to confirm.
7) VFD displays TEST COUNT=2, which means setting steps of testing, 20 steps are
permitted at most, here, they are 6 steps. Press 6 to set TEST COUNT=6, press Enter to
confirm.
8) Setting the mode of current procedure, press ▲▼to choose one mode in CONST
CURRENTCONST VOLTAGECONST POWERCONST RESIS. Here, the first step is
CC mode, press ▲▼to >CONST CURRENT, press Enter to confirm.
9) VFD shows SET 1=20.000A, which means setting the current procedure. Here, the first
sis 0.21A, press 0 + + 2 + 1 , then press Enter to confirm.
10) Setting whether the current procedure is short circuit or not, press ▲▼to choose
one in SHORT ON or SHORT OFF. Here the first step is SHORT OFF. Press SHORT OFF,
press Enter to confirm.
11) Select the test number of this step, press▲▼ to choose >READBACK V, Press
Enter to confirm.
12) VFD displays MIN 1=120.00V, which means setting the lower limit. Here, the first step
is 5.8V. Press 5 + + 8 , press Enter to confirm.
USER’S GUIDE 37
Share Link: GO URL