FRONT PANEL OPERATION
13) VFD displays MAX 1=120.00V, which requests setting the upper limit. Here, the first
step is 6.15V, press 6 + + 1 + 5 ,then press Enter to confirm.
14) VFD shows DELAY 1= 1.0〈S〉, which means setting the prolong time, that is how
long we can read the testing value after inputting the value we settled. To wait for testing
after the input stable, the range of prolong time is 0~25.5 seconds, when the setting is 25.5
seconds, automatic testing will be pause, only the user press Shift + Trigger ,the testing
will continue. Here, the first step is 1 second, press 1 and then press Enter to confirm.
15) Repeat 9)~14), and set the other process step by step as follows:
A. CONST CURRENT,0A,SHORT OFF, READ BACK V, 5.9V, 6.4V,1S
B. CONST VOLTAGE, 5V, SHORT OFF,READ BACK A, 0.205A, 0.245A,1S
C. CONST VOLTAGE, 3V, SHORT OFF, READ BACK A, 0.205A,0.245A,1S
D. CONST VOLTAGE, 2V,SHORT OFF, READ BACK A, 0.205A,0.245A, 1S
E. CONST CURRENT, 0A, SHORT ON, READ BACK A,0A,0.245A, 1S
16) VFD displays SHORT TEST FILE*, which requests saving the files edited to EEPROM,
automatically testing files and LIST files use a common memory area, and the area can
contain 8 groups of automatic testing files for next usage. Here, the testing file could be
saved in the first group, press 1 and press Enter to confirm.
17) When the edition of automatic testing file is finished, press Enter two times to back
the menu.
Fast Recall the Test File
1) Press Shift + Menu to enter into menu, VFD displays>CONFIG
2) Press▼ and move the menu to>LIST SET
3) Press Enter to enter into submenu, VFD displays>MODE SET
4) Press▼ and move the menu to>CALL TEST FILE
5) Press Enter and recall the file edited
Automatic Testing
After editing the automatic testing file, the automatic testing could be carried out by the
below steps.
1) Press Shift + l-set to enter into automatic testing, VFD displays the name of current
file: NAME: TEST FILE1
2) Press Enter to observe the voltage and current we inputted, and also can observe the
step number of testing.
3) Press Shift + Trigger and begin to operate, when the automatic testing is pause (the
current prolong time of one step is 25.5S), press Shift + Trigger also can make the testing
continue.
38 USER’S GUIDE